Composting Biodegradation Analyzer

Applications
Meets the testing requirements of GB/T 19277.1-2011.
Meets the testing requirements of ISO 14855.1.
Suitable for biodegradability research on bio-based materials and products.
Suitable for biodegradability testing under composting conditions.
Suitable for standard tests to measure carbon dioxide release under other simulated aerobic composting conditions.
Working Principle
This experimental system simulates the biodegradation of bio-based materials under aerobic composting conditions.
Under aerobic conditions, microorganisms use bio-based materials as a carbon source, consume oxygen, and produce carbon dioxide, water, and other inorganic substances. The concentration of the generated carbon dioxide is monitored in real time using a high-precision infrared sensor. By integrating the measured carbon dioxide concentration with gas flow rate and time, the total amount of carbon dioxide released during sample biodegradation can be calculated.
The biodegradation rate is determined by the percentage of the carbon dioxide released during sample biodegradation relative to the theoretical amount of carbon dioxide that could be released.
Technical Parameters
Operating Conditions
Item | Specification |
Power Supply | AC 220 V ±10%, 50 Hz, single-phase |
Operating Temperature | 15–30°C |
Relative Humidity | ≤80% |
Operating Duration | Suitable for long-term continuous operation |
Total Power | 4 kW |
Overall Dimensions (L × W × H) | 1500 × 900 × 900 mm |
Gross Weight | 180 kg |
Oxygen Supply System
Item | Specification |
Upstream CO₂ Removal | Available |
Gas Supply Method | Continuous aeration with a built-in gas supply pump |
Gas Pump Control | Nine independently controlled channels |
Flow Rate Control | PLC control with automatic adjustment according to the set value |
Flow Rate Control Accuracy | ±1% |
Flow Rate Measurement | Real-time measurement using electronic flow meters |
Flow Rate Measurement Range | 0.01–500 mL/min |
Number of Electronic Flow Meters | 9 |
Flow Control and Measurement | Operate independently |
Temperature Control System
Item | Specification |
Temperature Control Method | Air bath |
Set Temperature Range | Room temperature to 90°C |
Temperature Setting Increment | 0.1°C |
Temperature Control Accuracy | ≤±0.1°C |
Reactor System
Item | Specification |
Number of Channels | ≥9 |
Reactor Volume | ≥2000 mL |
Reactor Design | Double-layer structure |
Water Addition | Manual |
Stirring Method | Manual |
Reactor Bottle Material | High-borosilicate hard glass |
Gas Supply Method | Bottom gas supply |
Condensation and Dehumidification System
Item | Specification |
System Components | Low-temperature thermostatic bath, fully automatic condensation bottles, and dehumidification bottles |
Channel Configuration | Each channel is equipped with a condensation bottle and a dehumidification bottle |
Condensation Method | Water-bath condensation |
Condensation Temperature | As low as 5°C |
Drainage Method | Condensate is discharged automatically at preset intervals |
Dehumidification Method | Double-layer dehumidification bottle design |
Connection Method | Condensation bottle and dehumidification bottle connected in series |
Detection System
Item | Specification |
Detection Method | Infrared sensor for carbon dioxide and oxygen sensor for oxygen |
Number of Sensors | 1 carbon dioxide infrared sensor and 1 oxygen sensor |
Display Method | Simultaneous display of concentration and cumulative amount |
Carbon Dioxide Detection Range | 0–10,000 ppm |
Oxygen Detection Range | 0–21.0% |
Carbon Dioxide Resolution | 1 ppm |
Oxygen Resolution | 0.1% |
Data Acquisition Interval | 5 minutes |
Sensor Dehumidification Protection | Available |
Gas Analyzer Compatibility | Can be used in combination with a gas analyzer |
Software System
Item | Specification |
Software | Proprietary software control system with independently owned intellectual property rights |
Data Acquisition | Real-time data acquisition and curve generation |
Supported Operating Systems | Professional editions of Windows 7, Windows 8, Windows 10, and other compatible systems |
Data Storage | Real-time data storage during testing |
Cloud Platform | Remote data access and review |
Continuous Operation | Supports continuous operation for more than 6 months |
Software displays:
Concentration and cumulative amount data
Concentration variation curves

Standard Configuration
No. | Name | Quantity | Unit |
1 | Main Unit | 1 | Unit |
2 | Reactor Bottles | 9 | Pieces |
3 | CO₂ Removal Device | 1 | Set |
4 | Condensation Bottles | 9 | Pieces |
5 | Low-Temperature Thermostatic Bath | 1 | Set |
6 | Dehumidification Bottles | 9 | Pieces |
7 | Connecting Hoses | 1 | Set |
8 | Software | 1 | Set |
9 | Supplied Documentation | 1 | Set |
Experimental Example
The following example presents the biodegradability test results of a sample tested by an organization using an G9000 composting biodegradation analyzer by infrared method, with reference to the test method specified in GB/T 19277.1-2011.
The test lasted 70 days. As shown in Figure 1, the biodegradation rate of cellulose reached 81.88% within 45 days and 98.18% within 70 days. At this stage, the increase in biodegradation across the experimental groups had largely ceased, and the curves had leveled off.
The experimental design is presented in Table 1.
Table 1. Experimental Design
Item | 1 | 2 | 3 | 4 | 5 | 6 | 7 | 8 | 9 |
Inoculum | 300 g | 300 g | 300 g | 300 g | 300 g | 300 g | 300 g | 300 g | 300 g |
Cellulose | 30 g | ||||||||
Sample 1 | 30 g | 30 g | |||||||
Sample 2 | 30 g | 30 g | |||||||
Sample 3 | 30 g | 30 g |
Features
Fully meets the testing requirements of GB/T 19277.1.
Uses a solid culture and open oxygen supply design to simulate highly aerobic conditions.
Provides high testing accuracy and online measurement of carbon dioxide release.
Displays both carbon dioxide concentration and cumulative amount, with data exportable in graphical format.
Offers a wide detection range, with an upper limit of up to 10,000 ppm.
Features a low-noise design with nine independently controlled gas supply channels and an operating noise level below 45 dB.
Uses air-bath temperature control with a temperature adjustment accuracy of ±0.1°C.
Features a top-opening temperature control chamber with a sliding door for convenient operation.
Supports cloud-based testing, allowing online monitoring of data and curves.