Geomembrane Burst Strength Tester

This instrument is developed and designed in accordance with the national standards for the determination of the burst strength of geomembranes and the impermeability performance of geosynthetic materials specified in JTGE50, GB/T 19979.1, and GB 17642.
It is suitable for testing the hydrostatic pressure resistance of nonwoven composite geotextiles (one fabric layer with one membrane layer, or two fabric layers with one membrane layer) manufactured using various nonwoven geotextiles or other materials as the substrate and polyethylene, polyvinyl chloride, or other materials as the membrane, under specified water pressure conditions.
This instrument features simple operation, an attractive appearance, stable performance, and convenient use.
Technical Parameters
No. | Technical Parameter | Specification |
1 | Accuracy | 0.5% |
2 | Liquid Inflow Rate | 100 mL/min |
3 | Water Collector Diameter | φ30.5 mm |
4 | Pressure | 2.5 MPa |
5 | Specimen Size | φ55 mm |
6 | Water Supply Method | Automatic water supply by pressure pump |
7 | Voltage | 220 V |
8 | Pressure Adjustment Range | 0–2.5 MPa |
9 | Overall Dimensions | 700 × 750 × 1200 mm |
10 | Total Weight | 80 kg |
Installation and Use
1. Connect the power supply and turn the power switch. The power indicator will illuminate, and the LCD controller will start up.
2. After the LCD controller starts up, the controller will display the operation screen, as shown in the figure:
3. Click the Settings button to configure the operating parameters, as shown in the figure.
First, select Operation Settings to set the required test Pressure and Time, as shown in the figure.
4. After setting the required test pressure and time, click Back to return to the main screen.
5. Open the water tank filling cover and add purified water until the tank is full.
6. Open the pressure regulating valve to allow water to flow into the bottom plate of the high-pressure chamber until it just begins to overflow.
7. Place the specimen flat on the mesh inside the water collector, ensuring that there are no wrinkles. Allow excess water to overflow to ensure that there are no air bubbles inside the specimen clamping device. Cover the specimen with the perforated plate and clamp it evenly and securely.
Note: For specimens composed of textile material and a membrane layer, the membrane side should face the water. For composite materials with textile layers on both sides and a membrane layer in the middle, the textile material in the annular clamping area along the edge of the specimen facing the water may be carefully peeled away. Alternatively, an adhesive such as silicone sealant may be applied to the corresponding area to ensure that no water leaks through the clamped section.
8.After installing the specimen, open the valve and click the Test button, as shown in the figure.
Increase the water pressure inside the specimen clamping device to 0.1 MPa. If the approximate range of the specimen's hydrostatic pressure resistance can be estimated, the test may also be started directly at the lower limit of that range.
Maintain the above pressure for at least 1 hour and observe whether water seeps through the holes in the perforated plate. If no water seepage occurs, increase the pressure in increments of 0.1 MPa, maintaining each pressure level for at least 1 hour, until water seepage occurs. The appearance of water seepage indicates that the specimen has developed a leakage point or has ruptured.
Record the pressure at the preceding pressure level as the specimen's hydrostatic pressure resistance value, to the nearest 0.1 MPa.
9. When using the apparatus specified in GB/T 19979.2, water seepage shall be determined based on the seepage flow rate. If the seepage flow rate is extremely low under a certain hydraulic pressure differential, the specimen may be considered to have no water seepage. A rapid increase in the seepage flow rate indicates that the specimen has developed a leakage point or has failed, and the test may be terminated.
Note: If water droplets appear in the holes of the perforated plate, wipe them away and observe whether water continues to seep out. If no further seepage occurs, the water is considered to have overflowed from the specimen edges, and the test may continue. If water continues to seep out after wiping, the seepage is considered to be caused by water passing through the specimen, and the test may be terminated.
If it is only necessary to determine whether the specimen meets a specified hydrostatic pressure resistance value, the pressure may be increased directly to that value and maintained for at least 1 hour. If no water seeps out, the specimen is deemed to meet the requirements.
Measure the hydrostatic pressure resistance values of the remaining specimens according to the procedure described above. If the three measured values differ significantly (with the difference between the two lower values exceeding 50%), additional specimens shall be tested.
10. The lowest of the three measured hydrostatic pressure resistance values shall be taken as the hydrostatic pressure resistance value of the sample. If more than three values are measured, the average of the two lowest values shall be used. If only one value is significantly lower than the others and is more than 50% below the second-lowest value, that value shall be discarded.
11.Calculate the test results in accordance with the relevant standards.
After the Test
After the test is completed, open the water drain valve of the high-pressure chamber. Clean the instrument's surface, and wipe dry the pressure chamber cover, the bottom plate of the high-pressure chamber, and the perforated plate. Apply rust-preventive grease to these components.
If the equipment will not be used for an extended period, open the drain valve at the lower right of the main unit to completely drain the water from the internal system to prevent damage caused by freezing.