Horizontal Vertical Flame Tester Control System Operation Instructions
Horizontal Vertical Flame Tester control system operating instructions are suitable for components of electronic products and household appliances, parts box components, such as: wires and cables, insulating shells of household appliances, switch panels, printed circuit boards and insulating materials, etc. It is suitable for the research of lighting equipment, low-voltage electrical appliances, household appliances, machine tool electrical appliances, motors, power tools, electronic instruments, electrical instruments, information technology equipment, electrical affairs equipment, electrical connectors and accessories and other electrical and electronic products and their components. Production and quality inspection departments are also suitable for the industries of insulating materials, engineering plastics or other solid combustible materials.
The panel consists of a liquid crystal display module and three key switches
1. ———, burning display, set burning time and increase or decrease buttons;
2. Long press the "Auto key" to set the burning time, burning times and extinguishing time. Press the "manual key" to terminate the setting and exit.
3, ---, flameout time, set time and increase and decrease times;
4. "Start button" to start the test;
5. "Combustion form" selection: vertical, horizontal,
6, "burning mode" selection: manual, automatic
7. "Power key", turn on and off the main power
Manual experimental steps
1. Press the "power button", turn on the power, "burning mode" is manual, and the indicator light is on when it is off.
2. Select vertical/horizontal "burning method".
3. Press the "start button", ignite and burn, the burning indicator lights up, the burning timer starts, and the ignition lasts for 3 seconds.
4. Press the manual stop button to extinguish the flame, the indicator light will turn on, and the burning time can be read at this time.
5. If the burning time exceeds 60 seconds, please stop it manually.
6. During the test, the test can be stopped immediately and the flame can be extinguished by pressing the manual stop button or the power button.
Automatic method experimental steps
1. Press the "power button" to turn on the power, and select "burning mode": automatic.
2. Set the burning time, extinguishing time and burning times.
3. Select vertical/horizontal "burning method".
4. Press the "Start button".
5. Ignite and burn, the burning indicator lights up, the burning timer starts, and the ignition lasts for three seconds.
6. After the burning continues for the set time, the flame goes out, the burning indicator goes out, the burning timer is reset to 0, and the extinguishing indicator lights up, and the extinguishing timing starts.
7. After the extinguishing lasts for the set time, ignite and burn, the burning indicator lights up, the burning timing starts, and the ignition lasts for three seconds.
8. If the number of times set by the test is reached, it will end automatically, and the flameout indicator will be on, otherwise, repeat 5-7.
9. After the test, if you want to continue to do the test in the automatic mode, you need to repeat it, and then repeat 4.
10. During the test, the test can be stopped immediately and the flame can be extinguished by pressing the manual stop button or the power button.
Horizontal Vertical Flame Tester Maintenance and Maintenance:
1. The machine should be kept clean, and cleaning work should be done frequently
2. Pay special attention to safety when using this burner, pay attention to grounding, and avoid placing flammable objects around.
3. After the test, the gas should be turned off immediately.
4. The machine should be kept and tested by professionals.
2022-10-27 10:08